Hydrogen Test Enclosures
Designed to allow Hydrogen Studies to be safely conducted inside laboratories. These enclosures are equipped with ATEX Certified Ventilation Fans and an Automated Valve Panel in order to declassify the area for use with Autoclave Systems, Load Frames, and DCPD Testing Equipment.
More Info
-
Cortest Hydrogen Test Enclosures are designed to allow Hydrogen Studies to be safely conducted inside laboratories. These enclosures are equipped with ATEX Certified Ventilation Fans and an Automated Valve Panel in order to declassify the area for use with Autoclave Systems, Load Frames, and DCPD Testing Equipment.
All Hydrogen Test Enclosures are designed with safety in mind and come with detailed Risk Analyses and recommendations for adapting them to meet regional standards and regulations. Each enclosure is equipped with an Emergency Shutdown System that monitors Gas Sensors, Flow Sensors, Intrinsically Safe Pressure Transducers, and Flame Detectors. In the event that abnormal conditions are detected, the system will trigger an alarm and disable all non-essential electronics not designed for use in hazardous areas. Test Enclosures can be customized to include subsystems for specialized testing. Valve Panel designs are available for the addition of a Gas Booster System for testing at pressures higher than the supply pressure. Gas Sub-panels can be added for additional gases including Methane / Natural Gas for gas blending research.
Load Frames may be equipped with DCPD Testing Equipment for crack initiation or fatigue studies. Autoclave Systems may be equipped with gloveboxes and purifiers to enable specimen loading inside innerted environments. Systems can be supplied with Autoclaves that conform to international standards such as PED, CSA, CRN, DOSH, and the China Special Equipment Manufacture License.
-
MEDIA
Hydrogen (H2) Gas
Methane (CH4) Gas
Natural Gas
Inert GasRATINGS UP TO
4350 PSI (30.0 MPa)
572 °F (300 °C)
TEST METHODS
ASTM E399 ASTM E1647
ASTM E1681 ASTM E1820
ASTM G129 ASTM G142 -
Materials of Construction
-Wetted Components (Vessels, Pull Rods, and Grips):
-316 Stainless Steel, A286 Stainless Steel
-Non-Wetted Components (External Load Components):
-17-4 PH Stainless Steel
–Sealing
-PTFE, FFKM, PEEK, EPDM
Features
-Pressures Up to 30 MPa (4350 PSI)
-Temperatures Up to 300 °C (572 °F)
-Safety System equipped with Intrinsically Safe and Explosion Proof Sensors
-Gas Sensors, Pressure Transducers, Flow Switches, and Flame Detectors provide safety and early hazard detection
-Explosion-Proof Fan with Speeds up to 1200 CFM to prevent hazardous gasses from building up
Designed to work with existing Cortest equipment including
-
PLC Enclosure System
-Valve Panel Control
-Fan Control
-Pressure Loss Detection
-Gas, Flame, and Flow Sensing
-Emergency Shutdown
-Alarm System
PC Controlled Data Acquisition System
DCPD Capabilities (30A/20V Power Supply)
LabVIEW Based Software Tests
-Slow Strain Rate Test
-Constant Load Test
-Fatigue Test
-Creep Test
-
-Slow Strain Rate Testing – NACE TM 0198
-Constant Load Testing – NACE TM 0177
-High Pressure, High Temperature (HPHT) Stress Corrosion Cracking (SCC)
-Electrochemical Studies
-DCPD Testing – ASTM E647
-New Product Development/Testing
-Supercritical Steam